Testing the Cell for Measurement of Diffusivity in Liquids
Authors: Palatý Zdeněk | Žáková Alena
Year: 2003
Type of publication: článek v odborném periodiku
Name of source: Chemical and Biochemical Engineering Quarterly
Publisher name: Croatian Society of Chemical Engineering Technology
Place: Záhřeb
Page from-to: 135-138
Titles:
Language Name Abstract Keywords
cze Testing the Cell for Measurement of Diffusivity in Liquids For the measurement of diffusion coefficients in liquids a method has been suggested where the time dependence of the component concentration in a small volume of the solvent, which is in the contact with the solution inside the vertical bundle of capillaries plugged at the lower end, is monitored. This method has been tested using sulfuric acid, which diffusivity is published. It has been found that the calculated values of acid concentration are in good agreement with those obtained experimentally. The calculated values have been obtained by a numerical solution of Fick?s second law, the concentration dependence of diffusivity of sulfuric acid in the capillaries being considered. kyselina sírová;difúzivita;difuzní cela
eng Testing the Cell for Measurement of Diffusivity in Liquids For the measurement of diffusion coefficients in liquids a method has been suggested where the time dependence of the component concentration in a small volume of the solvent, which is in the contact with the solution inside the vertical bundle of capillaries plugged at the lower end, is monitored. This method has been tested using sulfuric acid, which diffusivity is published. It has been found that the calculated values of acid concentration are in good agreement with those obtained experimentally. The calculated values have been obtained by a numerical solution of Fick?s second law, the concentration dependence of diffusivity of sulfuric acid in the capillaries being considered. sulfuric acid;diffusivity;diffusion cell